You are to design a new multi-channel temperature data logging system for use in the hospital in which you are working. The specifications given for its design are:
To convert the resistance, R, of the thermistors to a proportional voltage a constant current source is used. However, the problem that you immediately encounter is that the thermistors have an exponential response curve which makes direct digitization over the entire temperature range of interest difficult. You decide to linearize the voltage, as a function of temperature, prior to digitization by taking the double logarithm of the voltages measured across the thermistors, (the result of which is shown in Fig. 1).
The process of taking the double logarithm of the analog signal is accomplished by the use of a pair of TL441AM logarithmic amplifiers (from Texas Instruments). The output voltage is then amplified and level shifted to give appropriate levels for the subsequent A/D converter. This being done, you collect the following data to see if your efforts have been justified:
Table 1: Thermistor Voltage and Temperature readings (after signal conditioning).Temperature, T (°C) | -40 | -20 | 0 | 20 | 40 | 60 | 80 | 100 |
Resistance, R (Ω) | 75773.0 | 21862.1 | 7352.90 | 2813.11 | 1199.72 | 560.34 | 282.64 | 152.78 |
Output from log-amplifiers, (V) | 4.579 | 3.970 | 3.369 | 2.775 | 2.185 | 1.594 | 0.999 | 0.399 |
a) First find some intermediate results, (letting temperature be represented by 'y' and voltage by 'x').
Now using these results one can determine estimates for the slope, m, and the y-intercept, b.
yielding the following equation for the best fit line,
b) First determine some further intermediate results,
SSE = Syy - Sxy2/Sxx ≅16800-(-500.3)2/14.8990775 ≅ 0.2961257
where n=8, is the number of data points, and
t α/2, ν = t 0.025, 6 ≅ 2.447 (from table)
Now,
and,